CALIBRATION:
1. Open daylight plate, and place 2-3 drops of distilled water on the main prism. Close the daylight plate so the water
spreads across the entire surface of the prism without air bubbles or dry spots. Allow the sample to temperature adjust
on the prism for approximately 30 seconds before going to step #2. (This allows the sample to adjust to the ambient
temperature of the refractometer)
2. Hold daylight plate in the direction of a light source and look into the eyepiece. You will see a circular field with
graduations down the center (you may have to focus the eyepiece to clearly see the graduations). The upper portion of
the field should be blue, while the lower portion should be white. (The pictures shown here and shown in step 3 & step
4 are only as reference; the specific scale is listed on the product.)
3. Look into the eyepiece and turn the calibration screw until the boundary between the upper blue field and the lower
white field meet exactly on the zero scale. The instrument is temperature compensated for use between 50°F and 86°F
(10°C to 30°C). For best results, calibrate when the temperature of the instrument is 68°F (20°C).
OPERATION:
NOTE: It is recommended to calibrate the instrument before
first use. The instrument is temperature compensated for use
between 50°F and 86°F (10°C to 30°C). If the instrument is
used outside of that temperature range, it is recommended
that it is calibrated before every use.
1. Lift the daylight plate off of the prism.
2. Place a few drops of the fluid sample on the clean, dry
prism.
3. Lower the daylight plate onto the prism. The fluid should
cover the prism without bubbles or dry spots. For best
results, wait 20 to 30 seconds for the temperature of the
fluid to match the temperature of the prism.
4. Hold the instrument so light goes into daylight plate.
5. Read the scale for the fluid sample at the line between
the blue area and white area. In the samples (at right),
the reading for Ethylene Glycol is -25°C (40.5% Ethylene
Glycol). The eye piece can be turned to bring the scale
into focus.
6. Clean the prism and daylight plate with a soft, clean cloth
dampened with distilled water. Make sure the instrument
is dry before replacing in its box.
For parts or service contact the service department: 1-888-825-6989
WARNING: This product can expose you to chemicals including lead, which is known to the State of California to cause cancer.
For more information go to www.P65Warnings.ca.gov
4
www.mastercool.com
-84 (70%)
60 63%
-70 (64%)
1.400
-60 (59%)
50 60%
40 57%
-50 (56%)
30 53%
1.350
-40 (52%)
20 49%
-30 (48%)
10 44%
1.300
-20 (44%)
5 41%
0 38%
-10 (39%)
-5 (36%)
1.250
5 34%
0 (33%)
10 30%
1.200
5 (29%)
15 25%
10 (25%)
1.150
15 (21%)
20 19%
20 (16%)
1.100
25 12%
25 (10%)
BATTERY
PROPYLENE
FLUID
ETHYLENE
GLYCOL
GLYCOL
32 0%
32 (0%)
F
-64.4 (70%)
51 63%
-56.6 (64%)
1.400
-51 (59%)
45.5 60%
40 57%
-45.5 (56%)
34.4 53%
1.350
-40 (52%)
28.8 49%
-34.4 (48%)
23.3 44%
1.300
-28.8 (44%)
20.5 41%
17.8 38%
-23.3 (39%)
1.250
-20.5 (36%)
15 34%
-17.8 (33%)
12.2 30%
1.200
-15 (29%)
9.4 25%
-12.2 (25%)
1.150
-9.4 (21%)
6.6 19%
-6.6 (16%)
1.100
3.9 12%
-3.9 (10%)
BATTERY
ETHYLENE
PROPYLENE
FLUID
GLYCOL
GLYCOL
0 0%
0 (0%)
C
Pos. Nr. Teilenummer Teilebeschreibung
1
43705-01
Tageslichtplatte
2
43705-02
Kalibrierschraube
3
43705-03
Fokuseinstellung
4
43705-04
Hauptprisma-Baugruppe
5
43705-05
Gummigriff
6
43705-06
Okular
EIGENSCHAFTEN:
• Bietet eine schnelle und einfache Möglichkeit zur Prüfung des Kühlmittel-/Frostschutzmittel-Gefrierpunktes
• Kompatibel mit Ethylen- oder Propylenglykol
• Misst Batteriesäure-spezifisches Gewicht
• Automatischer Temperaturausgleich
• Große, gut ablesbare Grafiken und Linse mit hoher Kontraststärke
• Einstellbarer Fokussucher
• Solider Metallgehäuseaufbau
• Enthält Aufbewahrungskoffer, Flüssigkeitspipette und Kalibrier-Schraubenzieher
• Nicht für Gebrauch mit OAT-Kühlmitteln oder DEF geeignet
SPEZIFIKATIONEN:
Wertebereich
Minimum Division:
Ethylenglykol: -84 - 32˚F
Ethylenglykol: 5˚F
Propylenglykol: -60 - 32˚F
Propylenglykol: 5˚F
Batterie: 1.100 - 1.400 Spezifisches Gewicht Batterie: 0.01 Spezifisches Gewicht Batterie: ±0.01 Spezifisches Gewicht
WARNUNGEN/WARTUNG:
1. Genaue Messung hängt von sorgfältiger Kalibrierung ab. Um genaue Ergebnisse zu erhalten, müssen Prisma und
Probe die gleiche Temperatur haben.
2. Setzen Sie das Messinstrument keinen feuchten Arbeitsbedingungen aus und tauchen Sie dieses nicht in Wasser ein.
Wird das Instrument diffus, so ist Wasser in das Gehäuse eingetreten. Kontaktieren Sie zur Reparatur Ihren Händler.
3. Messen Sie mit diesem Instrument keine abrasiven oder korrosiven Chemikalien. Diese können die
Prismabeschichtung beschädigen.
4. Reinigen Sie das Instrument zwischen jeder Messung mit einem weichen, feuchten Lappen. Unterlassung
regelmäßiger Reinigung des Prismas führt zu ungenauen Ergebnissen und Schäden an der Prismabeschichtung.
5. Hierbei handelt es sich um ein optisches Instrument. Dieses erfordert vorsichtige Handhabung und Aufbewahrung. Eine
Unterlassung kann zu Schäden an den optischen Komponenten und an seiner Grundstruktur führen. Bei vorsichtiger
Behandlung wird dieses Instrument jahrelangen, verlässlichen Dienst bieten.
Ein gutes
Zuviel RI-
Facettenbild
Flüssigkeit
www.mastercool.com
43705
REFRAKTOMETER
2. Kalibrierschraube
3. Fokuseinstellung
1. Tageslichtplatte
4. Hauptprisma-Baugruppe
5. Gummigriff
Accuracy:
Ethylenglykol: ±5˚F
Propylenglykol: ±5˚F
Nicht genug
RI-Flüssigkeit
Deutsch
6. Okular