en - Areas of application
A 865 installation
30
Installation
A 865 adapters are required to reprocess the reprocessable cannulae
from 3F and MF syringes used in dental applications. The adapters
are compatible for use with standard 3F and MF syringes made by
manufacturers such as KaVo, Luzzani and Sirona.
Screw the A 865 adapters into the connections on the injector man-
ifold.
When A 865 adapters are used, the wash water must be filtered.
The injector manifold must be fitted with an A 800 tubular filter. The
wash water consists of the inflowing water, the process chemicals
and the soiling.
The tubular filter is supplied with its own operating instructions.
Follow in particular the instructions on how to perform regular
cleaning of the tubular filter.