b) Following the injection, a drop of blood or fluid may
appear. Blood and/or fluid loss can be minimized by
immediately applying pressure to the injection site
for approximately 60 seconds with a cotton ball or
gauze. Apply adhesive bandage if needed or desired.
c) To ensure delivered dose to target tissue, refer to
Section 4) Give Injection for proper injection
technique. For a partial or incomplete injection,
refer to the medication or vaccine product labeling.
d) In the event of changes in performance of the
device, DO NOT USE, contact PharmaJet.
Cleaning, Disinfecting, and Storage
CAUTION: DO NOT autoclave.
CLEANING & DISINFECTING: Injector and Reset
Station may be cleaned by wiping excess debris
from the exposed surfaces using 70% Isopropyl
Alcohol.
STORAGE: To store Injector in Reset Station, insert
Injector and lift upward on the two sliding arms while
closing the lid. This action allows Injector to be stored
in the released state.
NOTE: The container and Injector are resistant to short
term duration exposure to water.
DISPOSAL: Dispose of used disposables in accordance
with local requirements.
Storage and Transport
Item
Injector
Reset Station
Syringes – 100 Bulk units
Universal Filling/Vial Adapters –
100 Single units
13mm Filling/Vial Adapters –
100 Single units
CAUTION: Use only PharmaJet components
with the Stratis
Temperature:
Reorder Information
Needle-free Injection System.
®
Operating
Temperature:
5°C
(41°F)
-40°C
(-40°F)
40°C
(104°F)
70°C
(158°F)
Catalog #
C501-10
C501-20
C501-34
C501-43
C501-46