LCD screen

OTHER INFORMATION

This product conforms to the provisions of the EC directive
93/42/EEC (Medical Device Directive). And also complies
with following standards (included but not limited):
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General requirements for safety
EMC standard:
EN 60601-1-2 Medical electrical equipment part 1-2: General requirements for
safety- Collateral standard: Electromagnetic compatibility- Requirements and tests
Performance standards:
EN 12470-5, Clinical thermometers - Part 5: Performance of infra-red ear
thermometers (with maximum device)
Follow instructions for use.
+40°C
(104°F)
The upper and lower limits of temperature.
+10°C
(50.0°F)
Keep dry.
Outside displayed temperature range I
Result is higher than 43.0 °C (109.4 °F).
➜ The user should rest in a steady room temperature for at
least 30 minutes. Then measure again.
Outside displayed temperature range II
Result is lower than 34.0 °C (93.2 °F).
➜ The user should rest in a steady room temperature for at
least 30 minutes. Then measure again.
Error symbol I
Disturbance occurred while measuring.
➜ Repeat the measuring steps again. If the error persists,
return the CTD504 to your local distributor.
Error symbol II
Memory error occurred while measuring.
➜ Repeat the measuring steps again. If the error persists,
return the CTD504 to your local distributor.
– Eng 7 –
Description & Solution