Zest Dental PrepStart H2O Manual De Operador página 4

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
3. Flip water toggle switch to"ON"position.
4. Depress foot pedal and allow disinfecting solution to flow through unit out the hand-piece tip into beaker/sink. Leave disinfectant solution in water lines for sufficient time to achieve disinfection.
Time period will depend on type of disinfectant used.
5. Flush remaining disinfectant solution through unit.
6. Purge unit with air, at least five (5) minutes to remove residual disinfectant from lines.
D. Handpiece Cleansing Procedure – External Surfaces (DO NOT STEAM STERILIZE HANDPIECE TUBING)
1. Place hand piece in autoclavable pouch and place in a gravity displacement autoclave and sterilize at 275º F for 15 minutes. Allow cooling before handling.
Do not utilize dry heat sterilizers or toaster ovens... i.e. Cox Sterilizer.
Symbol
RxOnly
The following symbols may appear on the product packaging or labeling.
Definition
Reference number
Use by
Caution: Federal law restricts this device to sale by
or on the order of a dentist.
European Mark of Conformity
Fig 1
Fig 2
DEFINITIONS OF SYMBOLS
Symbol
Serial number
Manufacturer
European Community Authorized Representative
Consult instructions for use
2
Definition