Description of your unit's symbols:
There are a number of technical markings on your unit. These can be
explained as follows:
Inside the battery compartment '+' indicates positive polarity and '-'
indicates negative polarity. The batteries' should be positioned as
indicated by three embossed outlines in the base of the compartment.
The unit and belt are made in China for Bio-Medical Research Ltd.,
Parkmore Business Park West, Galway, Ireland.
The unit requires three 1.5 volt DC batteries. DC is indicated by the
symbol:
Program
Parameters
14
0-70 milliamps, adjustable (up to 1K ohms load)
• Output Current
45, 50, 60, 70 and 75 pulses per second (PPS)
• Pulse Rate
Positive pulse width 200 microseconds / Negative pulse width 200 microseconds /
• Pulse Duration (Width)
Inter-phase interval 140 microseconds
• Mode # 1
Pulse rate 70 PPS / On time = 2.5 seconds / Off time = 2.5 seconds
Ramp Up and Down = 2 seconds / Timer = 20 minutes
• Mode # 2
Pulse rate 60 PPS / On time = 3.5 seconds / Off time = 3.5 seconds
Ramp Up and Down = 2 seconds / Timer = 25 minutes
• Mode # 3
Pulse rate 50 PPS / On time = 4.5 seconds / Off time = 4.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
• Mode # 4
Pulse rate 50 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 2 seconds / Timer = 40 minutes
• Mode # 5
Pulse rate 75 PPS / On time = 6.5 seconds / Off time = 6.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
• Mode # 6
Pulse rate 70 PPS / On time = 10 seconds / Off time = 10 seconds
Ramp Up and Down = 4 seconds / Timer = 30 minutes
• Mode # 7
Pulse rate 45 PPS / On time = 10 seconds / Off time = 10 seconds
Ramp Up and Down = 4 seconds / Timer = 30 minutes
• Timer
20, 25, 30 and 40 minutes depending on the mode.
• Compliance
Complies with UL60601-1
• Waveform
Symmetrical biphasic with zero net DC
• Maximum Charge Per Pulse
15 microcoulombs under normal conditions
• Power Source
3 x 1.5V AAA Batteries
• Tolerances
All electrical specifications are ± 10% into real 500-ohm load
English
This symbol means "Attention, consult the accompanying
!
documents".
This symbol means type BF equipment.
SN stands for "serial number". Inside the box is the serial number
specific to this unit. The letter preceding the number indicates the
year of manufacture, where "T" denotes 2014, "U" denotes 2015 etc.
The batch number of the belt is represented on the belt packaging,
by the number corresponding with the LOT symbol.
www.slendertone.com