Daily Disinfecting

CAUTION:
• Do not rinse or immerse the main unit in any
liquid.
• Do not wash or rinse any of the parts under
strong running water.
• Do not touch the mesh with your hand or any
other object.
• Do not use household bleach or any other
abrasive chemicals for disinfecting.
Daily disinfection is required for the medication
chamber and mouthpiece.
Step 1: Make a disinfecting solution
Make a vinegar solution (1 part white vinegar and
3 parts distilled water) OR mild detergent soap
(dishwashing soap in distilled water).
Step 2: Open medication chamber
Open the medication chamber and pour in a small
amount of disinfecting solution.
Step 3: Turn the device on
Nebulize for 1–2 minutes by pressing the START/STOP
button on the main unit.
While nebulizing, soak the mouthpiece in the
disinfecting solution for 10–15 minutes.
Step 4: Remove medication chamber
Once finished, remove the medication chamber from
the main unit. Discard any remaining disinfecting
solution from the medication chamber.
®
Mini-Mist
1.800.243.2636
25
Owner's Manual