Drip tray
Thank you for purchasing this piece of Bibby
Sterilin equipment. To get the best
performance from the equipment, and for
your own safety, please read these
instructions carefully before use.
This equipment is designed to operate under the
following conditions:-
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuation not exceeding 10%
❖ Overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
If the equipment is not used in the manner
described in this manual the protection provided
by the equipment may be impaired.
Mains on/off
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the
line supply corresponds to that shown on
the rating plate. This model requires a
supply rated at 220 - 240V, 50 Hz, ~, single
phase.
Power requirements
Model
SRT1& 2
There is an IEC socket at the rear of the
instrument for connection to the mains supply.
The unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a UK 3 pin plug and the other has a 2
pin "Shuko" plug for connection to the mains.
Choose the lead appropriate for your electrical
installation and discard the other. Should neither
lead be suitable, take the lead with the UK plug
and replace the plug with a suitable alternative.
Securing
screws
Wattage
50W