! CAUTION
To clean the boiler, only use Citrocal supplied by Nilo. Use 2 bottles of Citrocal for each
cleaning cycle and add water till the load pipe is full.
BED HEIGHT ADJUSTMENT FUNCTION
The bed lifts by pressing the SU (UP) key.
When the key is pressed, a consent signal is sent to the engine control unit for 15 seconds and the relevant
key is backlit blue for the same time. This happens even if the bed is already in its highest position.
The bed movement can be stopped by pressing the SU (UP) key again. Backlighting goes back to white.
Pressing the GIÙ (DOWN) key reverse the bed movement (and the associated actions on backlighting and
consent signals to the engine control unit).
The bed lowering function is similar.
The bed movement is not disabled during the boiler cleaning cycle.

WARNING SIGNALS

The following warning signals are envisaged
Warning signal type
Temperature sensor
Boiler failed to load
Boiler failed to unload
Bed temperature too
high
Pict.12 Alarms
BED PROGRAMMING
Description
Sensor open or
shortcircuited
The boiler float does
not indicate full level
after the load ELV was
open for TLOAD_C
time
The boiler float
indicates full level
after the load ELV was
open for TUNLOAD_C
time
The sensor measures
a temperature over
60°C
Message displayed
Err 1
Err 2
Err 3
Err 4
Edition
WET TABLE
001
Possible causes
Temperature sensor
failure
Tap closed
ELV load failure
Float failure
Control unit failure
Drain closed
ELV load failure
Float failure
Control unit failure
Resistor keeps giving
off heat. Board relay
glued
11
/