SP- SOFTWARE
SYSTEM REQUIREMENTS
SP- software is designed to run under Windows. Suitable versions are Windows
2000, Windows XP, Windows Vista and Windows 7.
The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
INSTALLATION
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double
click on Setup.exe)
3. Follow the on-screen instructions
NOTE Do not connect the I-Tec USB Sensor to the PC before the software is
installed
CONNECTING THE I-TEC USB SENSOR TO A PC
1. Connect the circular connector on the USB cable provided to the I-Tec USB sensor
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the SP- desktop icon
NOTE Do not disconnect the USB cable before closing down the SP-
program

CONFIGURATION

Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
Setting Output Processing
The temperature range for the 4 to 20 mA output, the averaging period and the
peak/valley hold processing can be set by using the main program menu to select:
[Setup] → [Output Processing]
The temperatures corresponding to 4 and 20 mA can then be set (minimum
–40°C,
span 1000°C).
To minimise the the effects of temperature fluctuations, noise etc on the output
signal, the averaging period can be set between 0 and 60 seconds.
If required, Hold Processing can be applied by setting hold mode to Peak Hold or
Valley Hold and setting a hold period between 0 and 1200 seconds.
Setting Target Emissivity
To set an Emissivity value for the I-Tec USB Sensor, use the main program menu
to select:
[Setup] → [Emissivity]
Emissivity can then be set to a value between 0.1 and 1.0, or selected from a
pre-
maximum 1000°C, minimum span 100°C, maximum
set list of materials by selecting [Data].
7