INSTRUCTIONS FOR USE
For single patient use (one person) only
INCLUDES INSTRUCTIONS FOR THE PARI LC
IMPORTANT: READ INSTRUCTIONS AND SAFETY PRECAUTIONS BEFORE USE
READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE. USE THE PRODUCT ONLY FOR ITS INTENDED USE
AS DESCRIBED IN THESE INSTRUCTIONS FOR USE. IMPROPER USE CAN RESULT IN SERIOUS OR FATAL ILLNESS/
INJURY, IMPROPER TREATMENT OR PROPERTY DAMAGE.
REMEMBER
REPLACE YOUR REUSABLE NEBULIZER CUP EVERY SIX MONTHS
PLUS NEBULIZER
®

Publicidad

Tabla de contenido
loading

Resumen de contenidos para Pari Vios PRO

  • Página 1 INSTRUCTIONS FOR USE For single patient use (one person) only INCLUDES INSTRUCTIONS FOR THE PARI LC PLUS NEBULIZER ® IMPORTANT: READ INSTRUCTIONS AND SAFETY PRECAUTIONS BEFORE USE READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE. USE THE PRODUCT ONLY FOR ITS INTENDED USE AS DESCRIBED IN THESE INSTRUCTIONS FOR USE.
  • Página 2 6. Patients under the age of 4 or any patients who have special needs should use a facemask instead of the valved mouthpiece. 7. Before use, check PARI Reusable Nebulizer for proper assembly. All parts should be seated firmly in place. SAVE THESE INSTRUCTIONS...
  • Página 3 INDICATIONS FOR USE: VIOS AEROSOL DELIVERY SYSTEM ® The Vios PRO is a light weight compressor for use with PARI Nebulizers. The Vios PRO Compressor is intended for ® ® adult and pediatric patients for use in hospital, clinic, or home environments.
  • Página 4: Getting Started

    C. GETTING STARTED Check your Vios PRO package to ensure you have the ® following items. It is important to have a second PARI Reusable Nebulizer on hand at all times. Air Filter Air Outlet Power Switch PARI Reusable Nebulizer Holder...
  • Página 5 Firmly press the Vios PRO. Please follow step by step. ® other end of the tubing to bottom of PARI reusable nebulizer (Fig. D). Clean/Disinfect Nebulizer Cup: Refer to Sections F & G to properly clean and/or disinfect your nebulizer prior...
  • Página 6 If activated at the time of first using rack of the dishwasher. Remove and dry the parts the PARI reusable nebulizer then it will serve as a reminder that the PARI reusable nebulizer can be replaced.   ...
  • Página 7: Maintenance & Repair

    10 minutes. compressor parts and could lead to malfunction. In the Hospitals and clinics can sterilize the PARI Reusable event fluid enters the unit, please contact PARI Patient Nebulizer parts (except mask and tubing) by Connection at 1.800.FAST.NEB (327.8632).
  • Página 8: Explanation Of Symbols

    NOTE: WHEN SWITCH IS IN THE “OFF” POSITION AND THE COMPRESSOR IS NOT RUNNING, THERE WILL STILL BE ELECTRICAL POTENTIAL INSIDE THE COMPRESSOR HOUSING. TO REMOVE POWER FROM THE COMPRESSOR, THE POWER PLUG MUST BE REMOVED FROM THE POWER SOURCE. EN-7 Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOSPRO...
  • Página 9 PLUS REUSABLE NEBULIZER LIMITED WARRANTY ® PARI reusable nebulizers include a 6 month limited warranty for home care use. If you have any problems and need to return your PARI reusable nebulizer, contanct PARI Patient Connection at 1.800.FAST.NEB (327.8632). Please be sure to clean and disinfect your PARI reusable nebulizer and place it in a plastic bag before returning.
  • Página 10: Guidance And Manufacturer's Declaration -Electromagnetic Immunity

    3 A / m 3 A / m of a typical location in a typical commercial or hospital IEC 61000-4-8 environment. NOTE U is the a.c. mains voltage prior to application of the test level. EN-9 Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOSPRO...
  • Página 11: Guidance And Manufacturer's Declaration -Electromagnetic Emissions

    If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Vios Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V ] V/m. Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOSPRO EN-10...
  • Página 12: Estas Instrucciones Se Aplican A La Nebulizador Lc

    INSTRUCCIONES DE USO Exclusivamente para uso individual (un paciente) ESTAS INSTRUCCIONES SE APLICAN A LA NEBULIZADOR LC PLUS ® IMPORTANTE: LEA LAS INSTRUCCIONES Y PRECAUCIONES DE SEGURIDAD ANTES DE USAR EL DISPOSITIVO. LEA TODAS LAS ADVERTENCIAS E INSTRUCCIONES ANTES DE UTILIZAR EL EQUIPO. USE EL PRODUCTO SÓLO PARA EL PROPÓSITO PARA EL QUE FUE DISEÑADO QUE SE DESCRIBE EN ESTAS INSTRUCCIONES DE USO.
  • Página 13: Precauciones De Seguridad - Lc ® Plus

    7. Antes de usarlo, verifique que el nebulizador reutilizable REFERENCIA FUTURA PARI esté armado correctamente. Todas las piezas deben estar perfectamente ajustadas en su lugar. El uso de un nebulizador armado en forma incorrecta puede mermar...
  • Página 14: Precauciones De Seguridad - Vios

    PRO AEROSOL DELIVERY SYSTEM ® El Vios ® PRO es un compresor liviano en peso para el uso con Nebulizadores de PARI. El Compresor Vios ® PRO está diseñado para pacientes adultos y pediátricos para uso en hospitales, clínicas o en entornos domésticos.
  • Página 15: Introducción

    Salida de aire Timestrip ® (Ubicación sugerida de ® Timestrip) Interruptor eléctrico Sujetador de el nebulizador reutilizable PARI Nebulizador reutilizable PARI (se incluyen 2) Tapa de la válvula de inspiración Inserto del nebulizador Semicírculo del nebulizador Anillos de alineación Salida del nebulizador Boquilla con válvula de expiración...
  • Página 16: Instalación Del Vios

    Conecte la entubación a la taza del nebulizador: Presione PRO. Sígalos uno por uno. con firmeza el otro extremo de la entubación a la parte inferior del nebulizador reutilizable PARI (Fig. D). Limpie/Desinfecte la Taza del Nebulizador: Consulte sección F-G para limpiar y/o desinfectar Inicie el tratamiento: Siéntese en una posición relajada y...
  • Página 17: Limpieza De Su Nebulizador Reutilizable Pari

    Extraiga y seque las piezas después solo registra el transcurso del tiempo después de que se ha activado. Si la activa al momento del primer uso del nebulizador reutilizable PARI, entonces le servirá como de que finalice el ciclo de la lavadora de platos.
  • Página 18: Desinfección Su Nebulizador Reutilizable Pari

    En caso de agua durante 10 minutos. que entre líquido a la unidad, comuníquese con PARI Patient Los hospitales y clínicas pueden esterilizar las Connection al 1.800.FAST.NEB (327.8632).
  • Página 19: Especificaciones

    NOTA: CUANDO EL INTERRUPTOR ESTÁ EN LA POSICIÓN “OFF” Y EL COMPRESOR NO ESTÁ FUNCIONANDO, IGUAL EXISTE POTENCIAL ELÉCTRICO DENTRO DE LA CARCASA DEL COMPRESOR. PARA DESENERGIZAR EL COMPRESOR, DEBE DESCONECTAR EL ENCHUFE ELÉCTRICO DE LA FUENTE DE ENERGÍA. ES-7 ¿Tiene preguntas? 1.800.FAST.NEB (327.8632) | PARI.com/VIOSPRO...
  • Página 20: Información De Garantía Y De Devolución

    ® compra, PARI extenderá su garantía limitada estándar en un (1) año, por un período total de garantía de seis (6) años a partir de la fecha de compra. Puede presentar un comprobante de compra de una de las siguientes maneras: 1) Proporcionar una copia del recibo o comprobante de compra o, 2) Identificar la fecha de compra y la empresa a la que efectuó...