Product Specifications:
Power supply
Frequency
Power consumption
Power Output
Pulse width
Trouble shooting: Causes and Solutions
Status of the product
No LCD display after the batteries
are installed.
No LCD display after power cord
was connected
No feeling but the LCD display is
in working
Stimulation becomes weak or
tingling
It stops output before the 30
minutes are up
Treated surface flushing red/
flushing, or tingling
4.5V, 3 x AAA
1 — 100HZ
Max. 540mw max.
Min. 25uA, 1.25mW
Max. 100uA , 20 mW (into 2K 0hm load)
20—220 µS
Cause
Battery contacts may need cleaning.
Batteries are low or polarity incorrect
The contact is bad between the batteries
and the spring
Batteries are low
The power is not connected
There is no power
You are using a non original power cord
Plug inserted in the jack firmly
Is the lead and the electrode connected
well?
Is the electrode terminal stuck onto the
skin correctly?
Dirt on electrode terminal surface?
Oil on skin
Batteries used.
Skin allergy against electrode terminal
material
Solutions
Check and remove
Check the cause and remove or re-install
batteries
Expand the anode spring by using a suitable
tool
Replace batteries with new
Check again if plugs have been inserted cor-
rectly, if necessary use the travel adapter
Power cord might be broken or damaged,
please try battery operation
Please use the original supplied power cord
Check and correct connection between elec-
trode and chord, chord and jack.
Connect them correctly
Stick the electrode terminal firmly onto the
skin
Wipe clean or replace electrode terminal
Clean your skin
Replace batteries or connect the power cord.
Check skin allergy history, for slight allergy,
change position or shorten treatment time,
for severe allergy, stop treatment or try
desensitization therapy.
5