Cleaning Or Replacing Filters - Miele A 814 Instrucciones De Manejo

Alojamiento para puntas para air scaler y piezo scaler
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 39
en - Areas of application
Cleaning the A
800 tubular filter
Replacing FP 20
filter plates
38

Cleaning or replacing filters

The tubular filter must be cleaned at least once a week.
We recommend cleaning the tubular filter after every 10 programme
sequences.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
Borosilicate filter plates must be replaced approx. every 2 weeks or
after 20 cycles have been run.
To do this, unscrew the adapter.
Unscrew the top part from the bottom part.
Remove the silicone adapter  and the filter plate  from the bot-
tom part.
Rinse out the bottom part.
Insert a new FP 20 filter plate into the bottom part. The plate must
rest fully on the ring-shaped support of the bottom part.
 Damaged or used filter plates must not be inserted under any
circumstances.
Insert the silicone adapter  into the bottom part.
Screw the top part and bottom part together.
Screw the adapter back onto the injector manifold.
Do not insert a filter plate into the holder if the injector manifold has
been fitted with an A 800 tubular filter.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

A 814/1

Tabla de contenido