10. EL GENER ADOR DE AIRE CALIENTE
Immunity test
IEC 60601
test level
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Compliance
Electromagnetic
level
environment – guidance
Where P is the maximum
output power rating of the
transmitter in watts [W]
according to the
transmitter manufacturer
and d is the recommend-
ed separation distance in
meters [m].
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey¹), should be less
than the compliance level
in each frequency
range²).
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
54
10. EL GENER ADOR DE AIRE CALIENTE
¹) Field strengths from fixed transmit-
ters, such as base stations for radio
(cellular/cordless) telephones and
land mobile radios, amateur radio,
AM and FM radio broadcast and TV
broadcast cannot be predicted theo-
retically with accuracy. To assess the
electromagnetic environment due to
fixed RF transmitters, an electromag-
netic site survey should be considered.
If the measured field strength in the lo-
cation in which the VOLCANO MEDIC
Vaporization System is used exceeds
Recommended Separation Distances between portable and mobile
RF Communications Equipment and the VOLCANO MEDIC Hot Air Generator
The VOLCANO MEDIC Hot Air Generator is intended for use in an electromag-
netic environment in which radiated RF disturbances are controlled. The customer
or the user of the VOLCANO MEDIC Hot Air Generator can help prevent electro-
magnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the VOLCANO MEDIC
Hot Air Generator as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
Separation distance according to frequency
output power of
transmitter
150 kHz – 80 MHz
[W]
d = 1,2 √P
0,01
0,12
0,1
0,38
1
1,2
10
3,8
100
12
the applicable RF compliance level
above, the VOLCANO MEDIC Vapori-
zation System should be observed to
verify normal operation. If abnormal
performance is observed, additional
measures may be necessary, such
as re-orienting or relocating the VOL-
CANO MEDIC Vaporization System.
²) Over the frequency range 150 kHz
to 80 MHz, field strengths should be
less than 3 V/m.
of transmitter [m]
80 MHz – 800 MHz
800 MHz – 2,5
d = 1,2 √P
d = 2,3 √P
0,12
0,23
0,38
0,73
1,2
2,3
3,8
7,3
12
23
55