Redox (O.R.P.) Calibration Menu
PROG
Config.
Calibr.
Probe
You can choose the automatic or manual mode.
-
Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; press
and the procedure is completed.
-
Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the value 465mV flashing, insert the probe into my solution, press
solution, then press
Alarms
Dislay
Alarm LED on
"Lev" icon flashing
Alarm LED on
"Lev" and "Stop" icons flashing
Alarm LED on
"Flw" icon flashing
"OFA" icon flashing
"Stop" icon flashing
"Alm" icon flashing
"Cal" icon flashing
EM00136321
+
Calibr.
Auto.
Calibr.
Manual
to confirm and start the calibration procedure as described previously.
End level alarm, without pump
operation interruption.
End
level
operation interruption.
Flow alarm activated, the pump has
not received from the flow sensor
the programmed number of signals.
O.F.A. alarm
The value read by the probe is out
of the alarm band set
Probe not calibrated alarm
Auto Cal
475mV
475mV
100% ok
Man Cal
475mV
475mV
100% ok
to exit the calibration (after 4 seconds the pump will
Cause
alarm,
with
pump
rev.2.0
EN
WAIT
60 s.
Error
475mV
WAIT
60 s.
Error
475mV
to display the value of my
Remedy
Restore the liquid level.
Restore the liquid level.
Press the
key.
Press the
flashing of the "Stop" icon, press
the key again to restart the pump.
Check the "Alarm Band" parameter
for
the
correct
programming mode.
Perform
the
probe
procedure.
and the display
and the display
key to stop the
settings
in
calibration
13