Defender Series Floor Scale
1.2. Connecting an Indicator
When connecting the base to TD52 indicator, push the base connector onto the
indicator connector and turn the locking ring clockwise to lock it in place.
When connecting the base to an indicator that does not have the matching
connector, install the optional load cell cable adapter (P/N 30429931 sold
separately) to the indicator. Then attach the load cell cable to the cable adapter.
Note: As an alternate wiring method, cut off the connector, strip the wires and
attach them to the indicator using the wiring codes in Table 1.
The extension cable accessory (P/N: 30429931) cannot be used if the scale will
be used in a legal for trade or legally controlled application, because there is no
provision to seal the connection between the indicator and platform.
TABLE 1. FLOOR SCALE WIRING COLOR
Indicator Signal Name
EXE+
SEN+
SIG+
SIG–
SEN–
EXE–
CODE
Floor Scale Wire Color
Green
Blue
White
Red
Yellow
Black
EN-3