Maintenance And Cleaning; Accessories - IKA MS 1 Indicaciones De Seguridad

Tabla de contenido

Publicidad

MS 1.1
Standard attachment
MS 1.2
One-hand attachment
MS 1.3
Test tube attachment with
one-hand insert
MS 1.4
Microtiter plate attachment
If the shaker is screwed down (i.e. built into a system), please
note that dynamic forces can be generated during operation as a
result of unfavourable loading and a poor centre of gravity positi-
on. This could cause the surrounding area to vibrate or induce
uncontrollable vibrations in the table. If you notice that the appli-
ance is not running smoothly, reduce the speed until it runs smoo-
thly again.

Maintenance and cleaning

The shaker MS 1 and MS 2 are maintenance-free. It is sub-
ject only to the natural wear and tear of components and their sta-
tistical failure rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
Please send in equipment for repair only after it has been cleaned
and is free from any materials which may constitute a health
hazard. Please tell us if you have used any materials which are
dangerous or which may constitute a health hazard.
Use only cleansing agents which have been approved by IKA
to clean IKA devices.
MS1/2 122004
Max. speed
Touch-
Continuous
[rpm]
operation mode
operation mode
2500
yes
no
1400
yes
yes
1400
no
yes
800
no
yes
To remove use:
Dyes
isopropyl alcohol
Construction materials water containing tenside/ isopropyl alcohol
Cosmetics
water containing tenside/ isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA. Wear the proper protective gloves during cleaning of the
devices. Electrical devices may not be placed in the cleansing
agent for the purpose of cleaning.
Before using another than the recommended method for cleaning
or decontamination, the user must ascertain with the manufactu-
rer that this method does not destroy the instrument.

Accessories

Designation
Description
MS 1.1
For small tubes and vessels L001260
Standard attachment
upto diameter 30 mm
MS 1.2
Consisting of MS 1.3
One-hand attachment
and MS 1.21
MS 1.3
Attachment for inserts
Test tube attachment MS 1.21, MS 1.31, MS 1.32,
MS 1.33, MS 1.34
MS 1.4
For microtiter plate
Microtiter plate attachment
MS 1.21
For vessels with flat
One-hand insert
base
MS 1.31
For 14 vessels
Test tube insert
with diameter 8 mm
MS 1.32
For 6 vessels
Test tube insert
with diameter 10 mm
MS 1.33
For 4 vessels
Test tube insert
with diameter 15 mm
MS 1.34
For any number of
Test tube insert
bore holes
Id-No.
MS 1
MS 2
X
X
L001900
X
X
L001490
X
X
L001280
X
-
L001540
X
X
L001840
X
X
L001850
X
X
L001860
X
X
L001830
X
X
13