Seca 757 Manual De Instrucciones Y Garantía página 76

Ocultar thumbs Ver también para 757:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 167
Fault
... only the item "SYS" is
visible in the "rF" menu?
... only the "SYS" and
"Lrn" items are visible in
the "rF" menu?
... the "APrt" and "tiME"
menu items are not visi-
ble in the "rF" menu?
... when the menu is
called up, the item "rF" is
not displayed?
... the display "Er:X:11"
appears?
... the display "Er:X:12"
appears?
... the display "Er:X:16"
appears?
... when the Enter key
(send/print) is pressed,
the "Er:x:71" display
appears?
... when the Enter key
(send/print) is pressed,
the "Er:x:72" display
appears?
76 •
Cause/Remedy
The wireless module is disabled.
- Activate wireless module ("Activating the wire-
less module (SYS)" on page 69).
The wireless module is activated but no wireless
group is set up.
- Set up wireless group ("Setting up a wireless
group (Lrn)" on page 69).
No wireless printer is registered to the wireless
group.
- Use the "Lrn" menu item to log the wireless
printer into the wireless group ("Setting up a
wireless group (Lrn)" on page 69).
The wireless module of the scale is defective.
- Please notify maintenance service.
The scale has too high a load, or too great a load
is applied in one corner.
- Reduce load on scale or distribute weight
more evenly.
- Restart scale.
The scale has been switched on with too great a
load.
- Remove load.
- Restart scale.
Oscillations have been induced in the scale, pre-
venting determination of the zero point.
- Do not touch the trolley or table supporting
the scale.
- Restart scale.
Data transmission not possible, wireless module is
disabled.
- Activate wireless module ("Activating the wire-
less module (SYS)" on page 69).
Data transmission is not possible, no wireless
group set up.
- Set up wireless group ("Setting up a wireless
group (Lrn)" on page 69).

Publicidad

Tabla de contenido
loading

Tabla de contenido