Publicidad

RF Radiata
3 Veff da
EN 61000-4-
80MHz a
3
2,5GHz
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in
each frequency range. Interference may occur in the vicinity of equipment
marked with the following symbol:
IACER Srl
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
d = 1,2 ⋅√P from 150kHz to
80MHz
d = 1,2 ⋅√P from 80 MHz to 800
3 Veff da
MHz
80MHz a
d = 2,3 ⋅√P from 800 MHz to 2,5
2,5GHz
GHz where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
22 di 24
MNPG72 Rev. 02 del 01/03/15