Installation

Precautions

This chapter gives detailed instructions and important information to install the Tiger II scale successfully.

Environment

Before you install the scale, identify the best location for the equipment. The proper environment enhances its operation
and longevity. Keep in mind the following factors, which might have a negative influence on the scale's operation:
☞Vibration
Vibration diminishes the scale's ability to measure accurately. Electrical machinery such as conveyors and drill presses
can cause inaccurate and non-repeatable readings. The scale may also read inaccurately if it is not leveled properly.
☞Air current
Moving air can cause the scale to read an additional force (add. weight) and have the same effect as vibration.
☞Friction
A scale cannot measure accurately if an object is rubbing or pressing against the platter.

Unpacking and inspection

Please inspect the package as it is delivered by the carrier. If the shipping container is damaged, check for internal
damage and file a freight claim with the carrier if necessary. If the container is undamaged, open the box, remove the
scale and place it on a solid, flat surface. Please keep the packing material and shipping insert in case you need to
return the scale to METTLER TOLEDO. The Tiger II is a precision instrument and may be permanently damaged if not
shipped in factory approved packaging.
Package contents for all Tiger II units include:
☞ Tiger II scale
☞ platter ☞ AC-DC power adapter
☞ tower incl. fixation material (tower version only)

Setup

Open the box and pull up the scale, including the packing material, then remove the packing material.
Tower versions (e.g. RN10-CNT) only:
Plug in the tower connector into the socket on bottom of the scale. Fix the cover of the chassis
Set the unit on a stable, level surface. Level the scale by turning the adjustable feet on the bottom of the unit. When the
bubble indicator is in the centre of the circle, TIGER II is leveled correctly.
Bubble indicator:
Good
All four feet must touch the surface to make sure that TIGER II does not rock.
Remove the transportation lock (3kg scale only) by:
1) removing the 2 screws;
2) putting the 2 rubber mats back.
Put the platter on top of TIGER II.
Unpack the power adapter and plug it into the bottom of the scale. Plug the line cord into a properly grounded AC
power outlet.
Power up sequence and turn off scale
FIX
Press the power key
ON/OFF
scale performs a diagnostic test on its ROM and RAM, then proceeds to normal operating mode. The power-up
sequence is as follows:
All segments of the display characters are lighted. This verifies operation of all segments.
The scale displays country code, GEO value, the software part number and revision number.
The scale displays "------", and captures zero and is ready for normal operation.
Note: Before switching on the Tiger II scale, always make sure there is nothing on the platter.
FIX
Press the power key
ON/OFF
Tiger II
Wrong
to turn on the Scale. The Tiger II goes through a series of self-tests when it is turned on. The
until weigh display window shows "oFF" to turn off the scale.