FLAEM Suction pro Instrucciones Para El Uso página 10

Aspirador profesional médico quirúrgico portátil
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 30
• Do not expose the device to particularly extreme temperatures.
• Keep the device away from sources of heat, direct sunlight or excessively warm locations.
• Never obstruct the air inlets positioned on both sides of the device.
• Always operate the device on an unencumbered hard surface.
• Always unplug the power cable immediately after use.
• The device is equipped with two easy to inspect safety fuses in case of malfunction. Before doing so, unplug
the power cable. Repairs must be done only by authorised FLAEM personnel. Unauthorised repairs void the
warranty and may be hazardous for the user.
ATTENTION: Do not modify this device without authorisation from the manufacturer.
• The Manufacturer, the Vendor and the Importer shall be held responsible for safety, reliability and perfor-
mance only if: a) the device is used in compliance with the instructions for use b) the wiring where the device
is being used is in compliance with safety regulations and current laws.
• The Manufacturer must be contacted to communicate unexpected problems/events related to functioning.
• Ensure that the connection tubing and the canister lid have been carefully sealed in order to avoid loss of suc-
tion.
• Do not upset the canister while the device is working as the liquid may be aspirated into the device, damaging
the pump. Should this occur, immediately switch off the suction machine, then empty and clean the canister
(send it to an authorised FLAEM service centre).
• When the safety device (10) intervenes, suction is halted. Switch off the device, empty the canister (9) and
carry out the cleaning procedure.
• We recommend personal use of accessories, collection canisters and connection tubing to prevent the risk of
contagious infection.
• The aspirator tip and the manual vacuum flow control are single use sterilised products and must be replaced
after each use.
• Check the expiration date on the original aspirator tip packaging and make sure the sterile packaging is intact.
FLAEM NUOVA declines any responsibility for harm to the patient correlated to the deterioration of the afore-
mentioned sterile packaging due to third party handling during the original packaging of the entire device.
• The hydrophobic/antiviral and antibacterial filter supplied with Flaem aspirators must be strictly replaced
for each new patient, or if the said filter is saturated.
Should the aspirator be used directly on the patient, the filter for Single-patient use must be replaced
within a maximum period of TWO MONTHS, or if the said filter is saturated.
INSTRUCTIONS FOR USE
Before each use, the accessories must be carefully inspected to ensure the absence of dust, incrustation, clots or
liquid substances both inside the connecting tubing and in the canister and its respective lid. Furthermore, they
must be cleaned following the instructions rigorously as stated in the "CLEANING, SANITISATION, DISINFECTION,
STERILISATION" paragraph. We recommend personal use of the accessories, the collection canisters and the con-
necting tubing to prevent risk of contagious infection.
Instructions for suction on the patient:
1. Connect the accessories, referring to the "CONNECTION DIAGRAM" on the cover.
ATTENTION: in addition to being antibacterial/antiviral, the hydrophobic filter also acts as secondary line of
safety and stops any fluids that the primary safety device in the canister lid accidentally does not stop.
Follow the directions below for proper assembly:
SCRIPT:
Side to connect to the VACUUM
inlet (12) of the collection canister
SCRIPT:
aspiratore side
10
Side to connect to the AIR INLET
of the suction machine

Publicidad

Tabla de contenido
loading

Tabla de contenido