Syringe diagram

Option 1
Locking Collar
> > > > > > >
Option 2
Locking Collar
> > > > > > >

lntended Use and lndications for Use

SIBELMED Volume Calibration Syringes are designed for testing
accuracy and if it's necessary, to calibrate spirometers. These
Calibration Syringes are manually operated by actuating the
piston rod in and out of the cylinder assembly. The outlet port of
the Syringe assembly is attached to the spirometer transducer.
The Calibration Syringe, when actuated, displaces known volumes
of air into the flow or volume measurement device. The operator
compares the calibration syringe volume with the results obtained
from integrating the flow or reading the displaced volume of
the measurement device to determine its accuracy. Calibration
Syringes are used in pulmonary function test laboratories
(both hospitals and clinics), bio-medical engineering calibration
laboratories, private physician offices, universities, research
User's manual S3000
User's manual S3000
Piston
Zero
Piston
Zero
EN 14
EN14
Outler port
Outler port
512-300-MUM
• REV. 2.01
Adapter