4. OPERATING YOUR COMPRESSOR NEBULIZER
Note: Prior to initial operation, the nebulizer should be thoroughly cleaned.
1. Place your compressor nebulizer on a flat and stable surface. Be sure that
you can easily reach the controls when you are seated. 2. Open the clear cover
and remove the inside accessories. Important: Prior to initial operation, the
nebulizer should be thoroughly cleaned referring to the "Cleaning Procedures"
in the manual. 3. Gently twist the top part of the nebulizer counter-clockwise to
disassemble the nebulizer. 4. Fill the bottom section of the nebulizer with the
medication prescribed by your physician. Be sure that the cone is put inside the
bottom section. 5. Gently twist the top part clockwise to re-assemble the
nebulizer. Be sure the two sections fit well. 6. Attach one end of air tube to the
base of the nebulizer. 7. Attach the other end of air tube to the air tube connector
located on the front of the compressor. 8. Attach mouth piece or mask per your
choice to the top section of nebulizer. 9. Plug the power cord into an appropriate
electrical outlet. Make sure at this stage, the power switch is at "OFF" status.
10. Press power switch to begin your prescribed treatment.
Important: The compressor motor has a thermal protector which will shut off the
unit before the unit is overheated. When the thermal protector shuts the unit off,
please: a. Switch off the unit. b. Unplug the unit from the electrical outlet. c. Wait
30 minutes for the motor to cool down before another treatment. Make sure the
air openings are not obstructed. 11. When treatment is finished, shut off the unit
and unplug it from the electrical outlet.
5. CLEANING: It is recommended that the nebulizer, mouthpiece and mask are
thoroughly cleaned with hot water after each use and cleaned with a mild
detergent after the last treatment of the day. If your physician or respiratory
therapist specifies a different cleaning procedure, follow their instructions.
Rinsing (after each treatment): 1. Disconnect the air tube, nebulizer, mouthpiece
and mask. 2. Gently twist the nebulizer to open it. 3. Rinse the nebulizer,
mouthpiece and mask with water. 4. Dry them with clean soft towel or let it air
dry. 5. Reassemble the nebulizer when completely dry and put these parts in a
dry, sealed container.
Disinfections: Please follow the following steps to disinfect your nebulizer
unless otherwise specified by your physician. It is suggested that the unit is
disinfected afterthe last treatment of the day. 1. Use one part white vinegar with
three parts distilled water. Make sure mixed solution is enough to submerge the
04
nebulizer, mouthpiece and mask. 2. Complete Rinsing Steps 1-3. 3. Wash
nebulizer, mouthpiece and mask in warm water and a mild detergent. Then
wash them in hot tap water. 4. Submerge these parts in the vinegar and water
solution for thirty minutes. 5. Complete Rinsing Steps 3-5.
Cleaning the compressor: Filter change 1. Wipe daily with a damp cloth.
2. Do not use any powdered cleaners or soap pads, which may damage the
finish. 1. Do not use cotton or any other materials. Do not wash or clean the
filter. Use only 's filters. And do not operate without a filter. 2. Change the filter
every 30 days or when the filter turns gray. 3. Changing procedure: A. Remove
the filter cover B. Replace the used filter with a new one. C. Replace the filter
cover.
6. REPLACEMENT PARTS & OPTIONAL ACCESSORIES: Replacement parts are
available through your local retailer or they can be ordered by mail. Optional
accessories include adult mask and baby pacifier inhaler. The pacifier inhaler
allows baby to receive aerosol treatment easily through normal breath.
7. SPECIFICATIONS
Power AC:
Power Consumption:
Medication Capacity:
Particle Size:
MMAD:
Sound Level:
Average Nebulization Rate:
Compressor Pressure Range:
Operating Pressure Range:
Liter Flow Range:
Operating Temperature Range:
Operating Humidity Range:
Storage Temperature Range:
Storage Humidity Range:
Dimension (L x W x H):
Net weight / Gross weight:
110V/120V/220V/230V 50/60HZ
180 VA Max
10 ml
0.5 to 10 um
3 um
<55dB
>0.3ml/min(2% saline solution)
29 to 52 Psi / 200 to 360 Kpa / 2 to 3.6 bar
11.6 to 19 Psi / 80 to 130 Kpa / 0.8 to 1.3 bar
5.0~8.0 lpm
10°C to 40°C (50°F to 104°F)
10 to 90% RH
-25°C to 70°C (-13°F to 158°F)
10 to 95 % RH
185mm × 110mm × 180 mm
1.5kg / 1.8kg
05