Maintenance And Upkeep; Periodic Checks - ORDISI IG-65DC Manual De Uso Y Mantenimiento

Ocultar thumbs Ver también para IG-65DC:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Once mounted the ceiling plate, assemble the horizontal arm nº 7 to vertical
arm nº 8, join the electrical connectors nº 14 and nº 15 and fix by washers nº
16 and screws nº 17.
4.2.3. - LAMP INSTALLATION
The electrical installation of the room where you plan to use the equipment should be in accordance with the rules UNE
(CEI60364-710), as well as the Low Voltage Regulation. Electrical installation of the room must include a fuse and a switch
network for the simultaneous separation of the lamp with electric supply.
It is essential that the lamp is properly grounded. Connect the network to the terminal located in the ceiling plate. The
supported Voltage is 100-240V AC.
Once you make the connections to check the proper operation of equipment and mount the embellisher nº 9.
4.3. - TURNING ON THE LAMP
1. Turn on the switch of the room. A red light should appear in the keypad when the lamp is connected.
2. Turn on the lamp, using the keypad No.5, and adjust the light intensity with the + and – buttons. (See Fig. 3).
3. To change the position of the lamp, use the handle nº 1. Do these gently avoiding brusque movements.
4. The lamp is adjusted at factory. If it is necessary a readjustment, see point 6.2
Fig. 3 – Keypad
4.4.- RECOMENDATIONS FOR USE
Move the lamp gently, avoiding brusque movements.
If the lamp is not in use do not leave it turned on, since apart from wasting electricity you will reduce the useful life of the LED.
6.
- CLEANING
Before doing anything make sure the lamp is off.
The lamp must be cleaned regularly with a cloth dampened with water, taking care that water does not reach the inside of the
equipment via the ventilation grille. For persistent stains such as blood, disinfectants, etc. a little neutral soap can be used. Do
not use products containing acetone in any case. Dry the lamp well before use. If there is liquid inside you must wait until it
has evaporated.
The handle located on the cupola is sterilizable by steam media at a temperature of 121°C with an exposure time of 20
minutes.

6. - MAINTENANCE AND UPKEEP

6.1. - PERIODIC CHECKS

According to hospital procedures, the maintenance service should periodically carry out the following checks:
1. Check the general working order of the lamp.
2. Check the mobility of the arm and cupola. See point 6.2.
3. Check that the equipment is still earthed and is correctly connected.
4. Check its general state (of cleanliness, etc...).
A – On / Off button
B – On / Off Led indicator
C – Luminous intensity adjustment buttons
D – Luminous intensity Leds indicators
Página 8 de 11

Publicidad

Tabla de contenido
loading

Productos relacionados para ORDISI IG-65DC

Tabla de contenido