2.
KEY FUNCTIONS
2.1
2.2
2.3
2.4
2.5
8
ON/OFF Key
Press once to switch on the instrument. Repeat to
switch off.
Holding down the key when switching on will display the
maximum pressure range of the instrument in PSI.
Holding down the key when switching off will display all
segments as a test.
ZERO Key
Press to zero the current readings on the display
to nullify any offsets caused by large temperature
variations. The instrument will retain the zero value in
memory even after switch off, so that the zero reading
will be displayed when the instrument is switched on
again. The zero key can be used whenever offsets
occur. Please note that pressing the zero key has no
effect whilst the instrument is in hold or record modes.
SCALE Key
Press to display the pressure measurement in each of
the different scales available. The selected scale unit
is retained in memory even after switch off, so that the
selected unit can be displayed when the instrument is
switched on again.
HOLD Key
Press to hold the displayed reading and press again to
resume measurement. The scale key will operate while
the instrument is in hold mode so that the held reading
can be displayed in different scales.
FILTER Key
Press to filter out transient readings. In filter mode the
instrument will average the readings taken over the
last 4 seconds (16 readings) to provide a more stable
reading on the display. Press the key again to resume
normal measurement.