Trouble Shooting Guide; Specifications - Mangar Health Airflo Mk3 Instrucciones Para El Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
Compressor does not run when 'up' is
selected on the 2-way hand control.
Ensure that the air hose from the product
being inflated is connected to the Airflo.
Compressor keeps stopping when trying to
inflate the lifting aid.
Check that there are no kinks in the air hose.
Battery State Indicator light flashing 'Red/
Green'.
This happens when the current overload cut-
out has tripped. Leave for 3 seconds to self
reset.
Airflo
Weight:
Output pressure:
Battery:
Battery Life:
Lifting performance (from full battery): Bathing Cushion - 4 bathing cycles;
Surfer Bather - 10 bathing cycles; Handy Pillow Lift - 11 lifts; Sit-u-Up - 20 lifts; Leglifter - 8 lifts
Environmental protection:
(Airflo housing)
(Push-button hand control) IP67 (Totally protected against dust & immersion in water)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation.
Charger: Model number - UE10WCP1-150035SPA
Input:
Output:
Double insulated Class II equipment
Simple solutions for everyday independence

Trouble shooting guide

Specifications

4.6Kg (10lbs)
0.35 bar (5 psi)
1 x 12Vdc - 2.3Ah (Maintenance free Valve Regulated Sealed Lead Acid)
Approx 1000 charging cycles (2 - 5yrs Depending on usage)
IP51 (Protection against large dust deposits & vertical falling drops of water)
AC 100v - 240v, 50-60Hz, 500mA
DC 15v, 350mA
Battery State Indicator light flashing 'Red'.
This happens when the compressor's internal
temperature cut-out has tripped. Leave the
Airflo for 60-90 minutes to cool.
Compressor causes electrical interference
with another piece of electrical equipment.
Move either the compressor or other item
further away from each other until the
interference stops.
Airflo Mk3
7

Publicidad

Tabla de contenido
loading

Productos relacionados para Mangar Health Airflo Mk3

Tabla de contenido