successful.
Adjustment options
You can individually adjust the back section or the
thigh section by using the buttons (4-7).
All motors can be simultaneously put in position
«up» or «down» with the buttons (2) and (3).
To reset to the horizontal position
Press the "all motors down" button (3) to simul-
taneously move all sections to the horizontal po-
sition.
Care and Maintenance
Your uni 22_75 RF is a high-quality product that
has been designed to serve you reliably for many
years to come. For occasional cleaning and the
removal of dust, we recommend the use of a dr y
cloth.
Changing the batteries in the radio remote con-
trol
Remove the cover from the battery compartment
(10) on the back of the radio remote control. Place
three A A A batteries into the compartment, paying
attention to the +/- markings. Close the compart-
ment cover again.
Changing the batteries (11) in the uni 20_75RF
motor (batteries for emergency lowering)
Carefully remove the batteries from the batter y
clips. Place two 9V block batteries in the compart-
ment, paying attention to the +/- markings. The
batteries in the motor provide power in the case
of a power failure.
Troubleshooting
In case of malfunctions with your uni 20_75RF,
consult the following troubleshooting tips. If the
problem cannot be resolved, disconnect the uni
20_75RF from the mains outlet and contact your
Swissflex stockist.
Loss of power in the sitting position
In case of a power failure, a batter y pack (2 x
9V) that has not been connected is located in the
motor control which allows you to move the bed
to the flat sleeping position. Connect this battery
pack and use the "all motors down" button (3) to
move the bed to the horizontal position when there
is no mains power.
The uni 20_75RF does not work despite being
switched on and all components being connected
properly
Disconnect the mains supply unit (12) from the
14
Operating instructions
mains outlet. Wait a minute and then reconnect
it. If necessary, connect a bedside lamp to check
that mains voltage (240V) is present. If there is no
current, check the fuse.
Press the "all motors down" button (3) to reset all
adjustments. The control of the uni 20_75RF con-
tains microprocessors, so irregular power supply
can lead to malfunctions.
The radio remote control does not work (The
background light does not work)
Insert new batteries. Where applicable, remove
any obstacles between the radio remote control
and the motor control.
After a power failure
Before resuming use of the uni 20_75RF, press
and hold the "All motors down" button (3), until
all comfort areas have reached the horizontal po-
sition and the motors have stopped moving.
Emergency lowering
The emergency lowering function allows the bed's
sleeping surface to be moved to the sleeping position
in case of a power failure.
1. Connect the batteries for emergency lowering
(16).
2. Push the button for emergency lowering (14) on the
motor (back).
Repeated use of the emergency lowering function
puts a lot of strain on the batteries, in which case they
must be replaced.
Technical data
Supply voltage: 110-240V
Remote control power supply: 3x AAA Batteries
Power rating motors: Intermittent operation FROM
2 min / 18 min
Power rating user socket : 850 W
User socket fuse: 5x 20 mm T 4A
Thrust motors: 4x 3000 N
Power supply for emergency lowering: 2x 9V block bat-
tery
uni 20 / uni 21_95RF
Problem
Possible cause
Remote control or
No mains supply.
control system not
Manual switch or control system defective.
working.
Switched capacitor or mains isolation blank.
Control systems
Thermoswitch on the transformer may have
suddenly no longer
been triggered or may be defective.
travel / move.
Thermal cut-out in the transformer may have
been triggered or may be defective.
Internal fuse may have been triggered or may
be defective.
No mains supply.
Supply line (mains and / or manual switch /
auxiliary control system) interrupted.
The battery-operated
Block battery / block batteries flat.
electric reset function
Block battery/batteries not connected.
cannot be activated.
GB
GB
Solution
Restore mains connection.
Contact your supplier / stockist.
Press the button on the mains isolation plug.
Leave the control system in the rest position for 20-30 minutes.
Contact your supplier / stockist.
Contact your supplier / stockist.
Restore mains connection.
Check the supply line and restore contact if applicable.
Check and replace the block battery / block batteries if necessary.
Connect the block battery / block batteries.
Operating instructions
15