Cleaning
Cleaning of the compressor
Use a clean dry cloth and mild detergents to clean the outside of
the unit.
ATTENTION During cleaning, make sure the internal parts of the unit
are not in contact with liquids and the power cord is disconnected.
Cleaning and disinfection of the accessories
After each use, disassemble T-piece kit into 3 sections. Wash in warm
soapy water and rinse well. Wipe with a paper towel and allow to air dry.
Do not boil or autoclave angled mouthpieces, air tubes, or masks.
It is important that the cleaning instructions are carefully
followed to ensure optimal operation.
Replacement of Nebulizer Kit
RES091 Disposable Nebulizer Kit with Clear Tubing
(Included in NEB100 and NEB100-R)
The disposable nebulizer T-Piece kit should be replaced after a long
period of inactivity, or if it is obstructed by dry medication, dust,etc. With
normal use, the disposable nebulizer T-piece kit should last 1 month.
RES093 Reusable Nebulizer Kit with Green Tubing
(Included in NEB100-R)
The reusable nebulizer T-Piece kit should be replaced after a long period
of inactivity, or if it is obstructed by dry medication, dust, etc. With
normal use, the reusable nebulizer T-piece kit should last 6 months.
Replacement of the filter
The filter is to be replaced
after approximately 30
hours of use, or when it
appears dirty. Open the
filter holder to replace
the filter with a new
one. Use Sunset BF018
Nebulizer Filters.
Do not use the device
without the filter.
Maintenance and Repair
Never open the device. There are no serviceable parts within the unit.
The compressor does not require lubrication or maintenance.
Filter cover
BF018 Foam Nebulizer Filter