SCOPE BUDDY™ FLUSHING AID CONNECTION DIAGRAM
SCOPE BUDDY™
Flushing Aid
Outlet Port
SCOPE BUDDY™
Adapter for Disposable Tubing - Male
Inlet Port
SCOPE BUDDY
Adapter for Disposable Tubing - Female
Fluid Intake Line
NOTE: Refer to SCOPE BUDDY Flushing Aid User Manual for operating
instructions.
FLUSHING TIMES FOR
24 HOUR SCOPE BUDDY DISPOSABLE TUBING
When using the SCOPE BUDDY Flushing Aid and 24 Hour SCOPE BUDDY
Disposable Tubing, all OLYMPUS
endoscopes should be flushed using a
®
programmed time of 60 seconds only.

CONNECTION INSTRUCTIONS

SCOPE BUDDY™ Adapter for Disposable Tubing - Female
Align the female adapter coupling to the "IN" port on the SCOPE BUDDY™
Flushing Aid and push on until it snaps into place. This adapter is reusable
(does not need to be discarded daily) and is sold separately, order number:
SB-110-HU1005.
SCOPE BUDDY Adapter for Disposable Tubing - Male
Align the male adapter coupling to the "OUT" port on the SCOPE BUDDY
Flushing Aid and push on until it snaps into place. This adapter is reusable
(does not need to be discarded daily) and is sold separately, order number:
SB-110-HU1005.
SCOPE BUDDY™ Disposable Tubing Delivery Tube
Align the male luer on the Delivery Tube with the female Luer on the SCOPE
BUDDY Adapter for Disposable Tubing and screw clockwise until fully
engaged. This tubing is for 24 hour use. Discard daily.
Fluid Intake Line
Align the female luer on the Uptake Tube with the male Luer on the SCOPE
SCOPE BUDDY™
BUDDY Adapter for Disposable Tubing and screw clockwise until fully
Disposable Tubing
engaged. This tubing is for 24 hour use. Discard daily.
Delivery Tube
WARNINGS & CAUTIONS
Failure to verify fluid flow through the endoscope may result in an
endoscope which is not properly flushed or cleaned. Always verify all
channel connections and if fluid is exiting all endoscope ports including
the distal end.
For 24 hour use only. Do not reprocess and discard daily.
Read all instructions prior to use. Failure to follow these may result
in inadequate flushing of endoscopes. Refer to SCOPE BUDDY™
Flushing Aid User Manual for additional instructions.
Install all necessary waterproof caps, plugs and cleaning adapters to
endoscope prior to immersion.
Users operating a SCOPE BUDDY Flushing Aid and SCOPE BUDDY™
Disposable Tubing must be trained and competent in the understanding
of endoscope channel systems.
SCOPE BUDDY™ is a trademark of Medivators Inc.
All other company or product names referenced in this document are trademarks
of their respective owner.
ENGLISH