Guidelines and manufacturer's declaration - electromagnetic interference
The SFT 76 is intended for use in the ELECTROMAGNETIC ENVIRONMENT described
below. The customer or the user of the SFT 76 should ensure that it is used in this type of
environment.
immuniTY
iec 60601-
tests
TesT leVel
ELECTRO-
± 8 kV
STATIC
Contact discharge
DISCHARGE
±2 kV, ±4 kV,
(ESD) in ac-
±8 kV, ±15kV
cordance with
air discharge
IEC
61000-4-2
Fast transient
± 2 kV for power
electric distur-
cables ± 1 kV for
bances / bursts
input and output
in accordance
cables
with IEC
61000-4-4
Impulse volt-
± 1 kV voltage
ages/ surges
external
in accordance
conductor
with IEC
External
61000-4-5
conductor
± 2 kV voltage
external conduc-
tor earth
140 GB/MT
comPliance
elecTromaGneTic
leVel
enVironmenT -
Guidelines
± 8 kV
Floors should be made from
Contact
wood or concrete or be dec-
discharge
orated with ceramic tiles. If
±2 kV, ±4 kV,
the floor is covered with syn-
±8 kV, ±15kV
thetic material, the relative
air discharge
humidity should be at least
30%.
Not applicable
Not applicable