CONTROL DIAL SETTINGS

DIAL:

POWER OFF

SET RATE
RUN
HOLD
VOLUME FED
VOLUME CLEAR
NOTE: Pump activates flow only in RUN setting. Pump maintains memory in all
settings except POWER OFF and VOLUME CLEAR.
FUNCTION:
Stops all pump functions. Battery charges in all
control dial settings when pump is plugged into AC
power.
Allows flow rate to be set (5 to 300 mL/hr in 1-mL
increments) by pressing the touch pads (arrow
buttons). No pumping action occurs in this mode.
Only setting that activates pumping action. Visual
display shows flashing RUN and rate of flow in
mL/hr.
Stops pumping action and silences audio alarm
(if one has occurred) without erasing previous
commands or memory functions. HOLD is used
when correcting an alarm condition, or whenever
feeding is temporarily interrupted (such as when
connecting a new feeding container).
Stops pumping action and displays cumulative
volume fed in mL, since this value was last cleared.
(NOTE: If 9999 mL is fed, numerical value turns
to 0 and a new accumulation begins.) To clear
VOLUME FED, turn knob to VOLUME CLEAR and
wait for 4 beeps to sound.
Stops pumping action and automatically clears to
zero any accumulated numerical value of volume
fed to that point.