Guidance and manufacturer´s declaration – electromagnetic emission
The Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer
or the user of the Infrared Thermometer should assure that it is used in such an environment.
Emissions
Compliance
test RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Guidance and manufacturer's declaration – electromagnetic immunity –for all EQUIPMENT and
SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or
the user of the Infrared Thermometer should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
±8 kV
discharge (ESD)
contact
IEC 61000-4-2
±15 kV air
Power frequency
30 A/m
(50/60Hz) mag-
netic field IEC
61000-4-8
Electromagnetic environment - guidance
The Infrared Thermometer uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The Infrared Thermometer is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings .
Compli-
Electromagnetic environment - guidance
ance level
±8 kV
Floors should be wood, concrete or ceramic tile. If floors are covered
contact
with synthetic material, the relative humidity should be at least 30 %.
±15 kV air
30 A/m
Power frequency magnetic fields should be at levels characteristic of
a typical location in a typical commercial or hospital environment.