INSTRUCTION MANUAL CODE 80135 REV C Feb/2019
Seg:
Eos:
Lym:
Mono:
Stab:
Baso:
Met My:
Myelo:
At Ly:
My Bi:
Normo:
Tox. Gr.:
Reti:
Ery:
J
.P. SELECTA s.a.u.
Segmentados o Granulocitos.
Eosinofilos
Linfocitos
Monocitos
Bandas
Basofilos
MetaMielicocitos
MieloBlastos
Linfocitos Atípicos.
Mielococitos Binucleados
NormoBlastos
Granulocitos toxicos
Reticulocitos
Eritrocitos / EritoBlastos
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.com - www.grupo-selecta.com
(Could be modified without notice)
6 WORKING WITH COUNTER
Power-up the cell counter by the rear panel switch.
Select some of the counter programs:
LEUCO program:
The LEUCO program uses 14 keys. Eleven of them, when pressed,
increase the total counting and partial counting corresponding key.
The other three ones «11. Normo» «12.Tox Gr» and «14. Ery»
increase the value of the corresponding key but not the total one
and therefore they don't have influence in the moment to calculate
the percentage.
MYELO program:
In this program all the keys, when being pressed, increase the value
of the corresponding key and at the same time increase the toal
value.
RETI program:
In this program there are only two active keys:
«13. Reti» and «14. Ery».
Key «13. Reti» increases their own value in 1 when being pressed
and it doesn't act on the total.
Key «14. Ery» increases their own value in 50,100,150,200 or 250
according to the elected configuration by the user in the SETTINGS
MENU and, at the same time, it increases the total in this same
quantity.
The «count end» is 1000.
FREE program:
In the FREE program all the keys are active and the user can choose
from the screen FREE PROGRAM which keys increase the total
and which don't.
Abbreviations used in the keys:
Segmented or Granulocytes
Eosinophiles.
Lymphocytes.
Monocytes.
Stab
Basophiles
MetaMyelicocites
MieloBlastos
Atypical Lymphocytes
Binucleated myelicocytes
NormoBlastes
Toxical granulocytes
Reticulocytes
Erytrocytes / Erythroblasts
Tel 0034 937700877 Fax 0034 937702362
Pag.: 13