  6HWWLQJ LQWR RSHUDWLRQ
After performing the installation procedures, the appliance can be set into
operation (Fig. 2):
1) fit the required electrode on the handpiece as indicated in § 4.1;
2) select the required function using selector (4);
3) switch on the appliance by means of the main power switch (1);
4) select the required power level by means of potentiometer (5);
5) instruct the patient to grip the neutral electrode;
6) press the button on the handpiece or press the pedal control; this
action will start the audible signal and deliver power to the electrode
attached to the handpiece.
Note that the appliance will assume alarm status after 20 seconds of
continuous use; this condition is signalled by illumination of the red AL
indicator (2) and a simultaneous intermittent beeping to inform the
operator that the current operation must be terminated as rapidly as
possible without prejudice to the patient; to continue using the appliance
without the alarm condition it must be switched off for approximately 10
seconds (by setting the main power switch to OFF).
The appliance can also be utilized without the ground electrode; this
procedure is permitted only for performing epilation or for low power
operations (less than 1/3 of the maximum output power).
  1RUPDO GHDFWLYDWLRQ DQG HPHUJHQF\ GHDFWLYDWLRQ SURFHGXUHV
The appliance can be deactivated as follows:
-
by releasing the button on the handpiece or the pedal control;
-
by setting the main power switch to OFF;
-
by unplugging the power cable from the mains socket outlet.