Flowtron ACS900
Hanger Bracket
Instructions for Use
IV Pole Bracket
1
Wall Bracket
1
Cleaning Instructions
The following processes are recommended, but should be adapted to comply with the local or national guidelines
(Decontamination of Medical Devices) which may apply within the healthcare facility or the country of use. If you are
uncertain, you should seek advice from your local Infection Control Specialist.
1. Clean all exposed surfaces and remove any organic debris by wiping with a cloth moistened with a simple (neutral)
detergent and water.
2. ArjoHuntleigh recommends a chlorine-releasing agent, such as sodium hypochlorite, at a strength of 1,000 ppm
available chlorine (this may vary from 250 ppm to 10,000 ppm depending on local policy and contamination status).
Wipe all cleaned surfaces with the solution, then wipe using a cloth moistened with water and dry thoroughly.
Alcohol based disinfectants (strength 70%) may be used as an alternative.
If an alternative disinfectant is selected from the wide variety available, we recommend that suitability for use is
confirmed with the chemical supplier prior to use.
ArjoHuntleigh AB
Hans Michelsensgatan 10
211 20 Malmö, Sweden
www.arjohuntleigh.com
For Use Only With Flowtron
2
2
Made In USA for ArjoHuntleigh
ACS900
®
WARNING
Mount the Flowtron ACS900 only on
an IV pole that can accommodate
pumps weighing 10 lb (4.5 kg) or
greater.
WARNING
Do not directly mount over a bed /
patient.
CAUTION
Ensure correct hardware is used for
the intended wall.
Ensure it is mounted level onto a
stable surface.
526359
526366
414514INT1 Rev A • 02/2016