Output current
Max. open circuit voltage
Immunity to spurious voltage
interference
Earth resistance of auxiliary
rods
Battery status checking
Advanced Features
Soil Resistivity Computing
Data Output
Built-in printer
Power supply
Safety Class
E.M.C.
Electrostatic immunity
Electromagnetic irradiation
immunity
Environmental Protection
Operation temperature
Storage temperature
Humidity
The short-circuit current is limited to less than
:
20.0 mA rms
:
50 V
During the R measurement, it allows for the
:
presence of spurious voltage up to 7 V~ , with a
error < 10%.
In the R measurement it allows Raux up to 50 kΩ
:
with error < 30%
The battery charge status is verified under normal
:
using conditions.
Automated detection of abnormal conditions that
:
may cause excessive errors (low battery, too high
noise interference, too high test spikes resistance)
When performing soil resistivity measurements, the
:
operator informs to the TL-6e the distance between
spikes and the equipment automatically computes
soil resistivity using the Wenner full equation.
:
USB
For a printed register document of measured
:
values.
By means of an internal rechargeable battery, LFP
:
12 V 3 Ah.
In accordance with IEC 61010-1
:
In accordance with IEC 61326-1
:
:
In accordance with IEC 61000-4-2
:
In accordance with IEC 61000-4-3
:
IP-54 with closed lid.
:
14°F to 122°F (-10°C to 50°C)
:
-13°F to 149°F (-25°C to 65°C)
95% RH (without condensation)
:
57