Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
· Slightly change the position of the electrodes.
GB
the stimulation feels Weak or not at all
• Check if the batteries need to be replaced.
· Electrodes are too old and need replacement.
the Broken circuit symBol is shoWn on the display symBol
The broken circuit symbol indicates that the resistance is too high, or that a cable
is broken.
· A too high resistance can be caused by a bad connection between the electrodes
and your skin, or that the electrodes need to be replaced.
· A cable breakage can be checked by pressing the cable´s pins against one anoth-
er while increasing the amplitude for the corresponding channel to 11 mA. if the
amplitude now drops to 0.0 mA and
placed.
note! Never increase the amplitude above 20 mA when you check for cable breaks,
since this can damage the stimulator.
the stimulator is not WorkinG

10. technical data

CEFAR EASY is an electrical stimulator intended for Pain Relief and Massage.
it delivers a constant current of 60 mA through a load of 1000 Ω.
information related to electromaGnetic compatiBility (emc).
CEFAR EASY is designed to be used in typical domestic or commercial environments
and approved according to the EMC safety standard of EN 60601-1-2.
CEFAR EASY emits very low levels in the radio frequency (RF) interval. Therefore it is
not likely to cause any interference in nearby electronic equipment (radios, comput-
ers, telephones etc.).
CEFAR EASY is designed to withstand foreseeable disturbances originating from
electrostatic discharges, mains supply magnetic fields and radio frequency trans-
mitters (such as mobile telephones).
8
if the display to the left appears when you start the stimulator, it
means that the stimulator is broken and needs to be replaced. NoTE!
do not use the stimulator - contact your Cefar dealer.
starts flashing, the cable needs to be re-

Publicidad

Tabla de contenido
loading

Productos relacionados para cefar easy

Tabla de contenido