Text ID Tag Operation:
To create a new Text ID Tag
•
Enter Meter or AUTO menu.
•
Press button above "FIBER ID", then press F4.
•
Press - or + to select new alphanumeric character, then
press Toggle Centre to move to the next character.
•
Press F2 to save or F1 to cancel.
To use a stored Test ID Tag
•
Enter Meter or AUTO menu.
•
Press button above "FIBER ID".
•
Press - or + to find required Tag,
•
Press F2 to confirm or F1 to cancel selection.
To modify a stored Text ID Tag
•
Enter Meter or AUTO menu.
•
Press button above "FIBER ID", then - or + to select a stored
ID Tag.
•
Press F3 to enter edit mode.
•
Press - or + to select new alphanumeric character, then
press Toggle Centre to move to the next character.
•
Press F2 to confirm or F1 to cancel modification.
To delete a Text ID Tag
•
Enter Meter or AUTO menu.
•
Press button above "FIBER ID", then - or + to select a stored
ID Tag.
•
Press F3 to enter edit mode, then press F4 to delete.
•
Press F2 to confirm or F1 to cancel deletion.
ORL Zero Function:
•
Set up the stray optical condition.
•
At Home menu, press F2 then press F3 twice to enter Noise
Floor (NF) setting mode.
•
Press F2 and, NF offsets for all active wavelengths will be
set and saved automatically. The new NF values will be
displayed next to the selected wavelength.
User Calibration (UCAL):
•
At Home menu, press F2 to enter ORL mode.
•
Press - or + to select wavelength.
•
Press F3 to enter UCAL offset setting mode.
•
Press - or + to enter the required value.
•
Press F2 to save or F1 to cancel the new settings.
Reset User Calibration (UCAL) offset & Zero Function (Noise
Floor) to factory defaults:
•
At Home menu, press F2 to enter ORL mode.
•
Press F3 twice to enter Noise Floor setting mode.
•
[For instrument with only 1x Two-way port]: hold down F2
then Toggle Centre until it stops beeping.
•
[For instrument with 2x Two-way ports]:
-
For SM port: hold down F2 then "-" until it stops
beeping.