MANUAL DE INSTRUCCIONES CODIGO 80021 REV F
J
.P. SELECTA s.a.
Mayo/2012
Do not use the incubator to dry or for thermal treatment of products
which can give off vapours capable of making an explosive mixture.
Be sure that the voltage supply is the same as the one indicated on the
characteristics plate.
Do not use the equipment if it is not earthed.
If you change the plug bear in mind the following :
OPERATION
CONTROL PANEL:
1. Main switch / Timer selector (optional).
2. Mains indicator lamp (green).
4. Safety thermostat control knob.
5. Safety thermostat blockage device.
6. Overtemperature indicator lamp (red).
7. Timer knob (optional).
8. Indicator lamp of end of cycle (amber) (optional).
9. Thermostat control blockage device.
10. Thermostat control Knob.
11. Heater operation indicator lamp (amber).
12. Thermometer.
13. Internal light switch (optional).
NOTE: The panel shown includes timer and safety thermostat, which
are supplied as accessories.
STARTING UP
1. Press the mains switch (1), at the same time the green indicator
lamp (2) will light up.
2. Turn the temperature control knob (10) clockwise until the working
temperature coincides with the pointer. The amber heater operation
indicator lamp (11) lights up.
3. Block the temperature control knob (10) with the blockage lever (9)
to avoid accidentally moving the knob.
4. After a certain time, the heater operation indicator lamp (11) will start
operating intermittently. When these intermittences become uniform,
the temperature of the incubator is stabilized.