Guidance and manufacturer's declaration - electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Immunity Test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IE 61000-4-2
Electrical fast
±2 kV for power
Transient/burst
supply lines
IEC 61000-4-4
±1 kV for
input/output
lines
Surge
±1 kV differntial
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips, short
<5 % UT
interruptions and
(>95 % dip in
voltage variations on
UT)
power supply input
for 0,5 cycle
lines IEC 61000-4-11
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec
(50/60 Hz) magnetic
3A/m
field
IEC61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
IEC 60601
test level
Compliance Test
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle 40 %
UT
(60 % dip in UT)
for 5 cycles 70 %
UT
(30 % dip in UT)
for 25 cycles <5 %
UT
(>95 % dip in UT)
for 5 sec
3A/m
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typical
commercial or hospital environment. IF the user
of the EQUIPMENT requires continued operation
during power mains interruptions, it is
recommended thta the EQUIPMENT be powered
from an uninterruptible powered supply or a
battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
7