Vyaire MicroCO Manual De Funcionamiento página 67

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 145
WARNING:
The MicroCO should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary, the
MicroCO and the other equipment should be observed /monitored, to verify
normal operation in the configuration in which it will be used.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Compliance
Group 1
Group B
Not Applicable
(battery
powered)
Not Applicable
(battery
powered)
Electromagnetic Environment -
Guidance
The MicroCO uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
The MicroCO is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes
65

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para Vyaire MicroCO

Tabla de contenido