Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Smoke Check is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or the user of the Smoke Check should
ensure that it is used in such an environment
Immunity
IEC 60601
Test
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to
80 MHz
<5% U
T
(>95 % dip in U
)
T
For 0.5 cycle
40% U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5% U
T
(>95 % dip in U
)
T
For 5 s
3 A/m
Compliance
test level
Level
(Battery
Powered)and
No Cables
Not Applicable
(Battery
Powered)
3 A/m
Electromagnetic Enviroment -
Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Smoke Check, including
any cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation
distance (d)
d = 1.2√P
N/A
28
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Smoke Check requires
continued operation during
power mains interruptions,
it is recommended that the
Smoke Check be powered
from an uninterruptable
power supply or a battery.
If incorrect operation occurs,
it may be necessary to
position the Smoke Check
further from sources of
power frequency magnetic
fields or to install magnetic
shielding. The power
frequency magnetic field
should be measured in the
intended installation location
to assure that it is
sufficiently low.