STARTING UP
1. Press the start switch (1) and the screen (2) will light up at the
same time.
2. Press keys (5)/(4) until you select the desired temperature value in
the green display (the upper one).
3. Press key (6) if you want to select an operating time.
4. Press key (3) to start the cycle.
5. The heating symbol
6. Press the key (3) again to stop the incubation process. Note: dur-
ing operation, the temperature or time adjustment value can be
changed.
COMMENTS:
* Do not use the incubator at a lower temperature than ambient +5ºC,
because the thermostat might have difficulties in regulating correctly.
* When using the time function, the green display (the upper one) will
indicate the remaining time. If you want to see the target temperature
value, press key (6).
* Place the load in the incubator trying to leave spaces for the air
circulation. Do not load the incubator more than 70% of its useful
chamber volume.
ADJUSTING THE SAFETY THERMOSTAT:
From the factory, the thermostat is adjusted to its maximum tempe-
rature.
To adjust the safety thermostat temperature, do as follows:
1. With the incubator connected to the mains, turn the safety thermos-
tat screw (7) clockwise to the end of the scale, with the help of a
flat screwdriver.
2. Select the desired safety temperature using the control keys. This
temperature has to be minimum at 3ºC above the working tempe-
rature.
3. With the door closed, turn on the incubator with key (3).
4. When the incubator is stabilized at the above selected temperature,
turn the safety thermostat screw (7) slowly to the left until it clicks,
and the red indicator lamp lights up.
5. Stop the incubator and leave it to cool with the door open during
15 minutes.