Operating Instructions E-Drive
62 61 957 D3436.201.05.05.
7.2
Automated cleaning and disinfecting
The cleaning and disinfection equipment used must be approved by its
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ The instrument is conditioned with a cleaning and disinfection
device.
1. Check whether the instrument is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
For further conditioning, the instrument should be dry and free of
residue.
3. Blow the instrument out with max. 3 bar.
7 Conditioning
37