HEINE BETA 200 Manual Del Usuario página 43

Ocultar thumbs Ver también para BETA 200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 15
Appendix:
Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Emission test
RF emissions
CISPR11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Guidance and manufacturer declaration - Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD) acc. to
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Comment: U T is the a.c. supply voltage prior to application of the test level.
Compliance
Group 1
The device uses RF energy only for its internal function. Therefore,
RF-emission is very low and it is unlikely that any interference in
nearby electronic equipment.
Class B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Warning: This device is intended only for use by medical professio-
nals. This is a device of class A CISPR 11 in the domestic environ-
ment, this device may cause radio interference, so that it may be
necessary in this case, to take appropriate remedial measures,
as e.g. orientation, new arrangement or shielding of the device or
restrict the connection to the site.
not
applicable
Passed
IEC 60601 test level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and
output lines
± 1 kV voltage
phase – phase,
± 2 kV voltage
phase – earth
< 5% U T , (>95% dip in U T )
for 1/2 period
40% U T , (60% dip in U T )
for 5 periods
70% U T , (30% dip in U T )
for 25 periods
<5% U T , (>95% dip in U T )
for 5 seconds
3 A/m
med 1114
Electromagnetic environment – Guidelines
Compliance level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and
output lines
± 1 kV voltage
phase – phase
± 2 kV voltage
phase – earth
< 5% U T , (>95% dip in U T )
for 1/2 period
40% U T , (60% dip in U T )
for 5 periods
70% U T , (30% dip in U T )
for 25 periods
<5% U T , (>95% dip in U T )
for 5 seconds
3 A/m
2014-10-15
Electromagnetic environ-
ment – Guideline
Floors should be wood, con-
crete or covered with ceramic
tiles. If floors are covered with
synthetic material, the relative
humidity should be
at least 30 %.
The supply voltage quality
should be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the device
requires continued operation
during power mains interrup-
tions, it is recommended that
the device be powered by a
UPS (uninterruptible power
supply) or a battery.
Power frequency magnetic
fields should be at levels
characteristic of
a typical location in a typical
commercial or hospital
environment.
43/48

Publicidad

Tabla de contenido
loading

Productos relacionados para HEINE BETA 200

Tabla de contenido