IMPORTANT: Start by reading the Easypump C-bloc RA Pump Instructions For Use,
then continue with this document.
Easypump RA with PCA: Disposable and Ambulatory Elastomeric Infusion System
(refer to Figure 1 on Page 4):
➊ Fill Port Cap
➋ Fill Port
➌ Outer Shell
➍ Elastomeric Membrane
➎ Clamp
Easypump C-bloc RA WITH PCA DESCRIPTION
Easypump C-bloc RA with PCA incorporates a bolus device. The PCA delivers a
continuous infusion (basal) and allows fixed boluses to be delivered on demand by the
patient or health care provider.
NOTE: For models including Select-A-Flow, see important information in the Select-A-Flow
product insert for setting the basal flow rate.
CAUTIONS:
•
Do not underfill the pump. Underfilling the pump may significantly increase the flow rate.
•
Do not add unvented filter to end of set as this may impede or stop flow rate.
•
Do not remove the red holding tab until the tubing is completely primed. Up to a 5 ml
bolus of air may be delivered if not primed correctly.
WARNING: Total Flow Rate refers to bolus + basal (see Table 2). To reduce potential
adverse effects:
1. Medication dosing should be based on the Total Flow Rate.
2. Red tab must be removed before use. If red tab is not removed before use or breaks
while removing, bolus button will not work and flow rate may increase significantly
above the Total Flow Rate.
3. If the bolus button does not pop back up within 30 minutes after delivering a bolus,
then close the clamp. The button may be stuck and flow rate may increase significantly
above the Total Flow Rate.
Delivery Time Information for the Easypump C-bloc RA with PCA
Table 1
Nominal
Maximum
Fill Vol
Fill Vol
270 ml
335 ml
400 ml
550 ml
600 ml
750 ml
When filled to nominal volume, accuracy is ±15% (except bolus dose +10/-20%) of the labeled
rates when infusion is started 0-8 hours after fill when delivering normal saline at 72°F (22°C).
NOTE: For models including Select-A-Flow, see Select-A-Flow product insert for basal flow rate
accuracy information.
➏ Air-Eliminating Filter
➐ PCA Bolus Device
➑ Air-Eliminating Filter
➒ Tubing
➓ Distal End Cap
Table 2
Retained
Vol
< 9 ml
< 15 ml
< 22 ml
Bolus
Refill
Dose
Time
5 ml
30 min
5 ml
60 min
5
Total Flow Rate
10 ml/hr + Basal Rate
5 ml/hr + Basal Rate