Equipment Description; Installation; Connection To The Mains; Operation - Selecta DIGITERM S-150 Manual De Instrucciones

Termostato de inmersión
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
Level switch for
liquid detection
Heater
(only DIGITERM
models)
The level switch inhibits the thermostat operation
if no liquid level is detected
(only DIGITERM models)
Fig 7.1. Fixing clamp
J
.P. SELECTA s.a.u.
CODIGO 80297 REV D
5. Equipment Description
Immersion thermostats are widely used on laboratories for thermal treatment of samples
and bath's heating.
The equipment includes a circulation pump that can be used to improve the homogeneity
within a bath or to perform a closed liquid circulation circuit for heating.
The unit could work in a continuous or timed operation. Once the operating time has
elapsed, the unit stops (only DIGITERM models)

6. Installation

Unpack the equipment and visually check that the contents are in good conditions. Keep
the package for some days.
6.1 Liquid tank
A bucket or tank will be needed to operate the thermostat. See J.P. SELECTA, s.a.u.
products range or paragraph 2 of this manual.
Circulation
pump
Fix the thermostat in the tank with the fixing clamp.
Place the equipment on a flat level stable surface, suitable for the equipment weight.
Select a suitable fluid for working temperatures, water, silicone oil, antifreeze, etc.
To work in a comfortable way, please leave a free space of 20cm around the equipment.
6.2 Connection to the mains
Plug the equipment to a near socket suitable for the equipment power. Use the connec-
tion cable supplied.
For your safety, the equipment must be earthed.
Before plugging the machine, check that the voltage supply and the frequency match
the one indicated in the equipment plate.

7. Operation

All the equipment functions are set with the touch screen keys. Gently press with your
finger in the center of each key.
Use the fixing clamp (see Fig 7.1) in the rear part of the equipment.
FLUIDOS ACONSEJADOS SEGÚN LA TEMPERATURA DE UTILIZACIÓN DEL BAÑO
RECOMMENDED FLUIDS DEPENDING ON BATH TEMPERATURE
Alcohol
Anticong. coche/Car antifreezer
-40
-20
0
20
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Julio 2019
(Sujetas a modificaciones sin previo aviso)
Agua / Water
*Rhodorsil 47v50 / Rhodorsil 47v50
40
60
80
100
120
140
160
180
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 9
200

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido