Applied Standards; Error Codes - Salter TE-250-EU Instrucciones Y Garantía

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 41

APPLIED STANDARDS

This product conforms to the provisions of the EC directive MDD(93/ 42/ EEC). The
following standards apply to design and/or manufacture of the products:
• ASTM E1965-98
Standard Specification for Infrared Thermometers for Intermittent Determination of
Patient Temperature.
• ISO 14971
Medical devices-Application of risk management to medical devices. The
Classification according to IEC/EN60601-1 sub-clauses 5:

ERROR CODES

When a malfunction or incorrect temperature measurement occurs, an error message will
appear as described below.
LCD Display
Cause
The temperature measured
is higher than
Human thermometer mode:
43°C (109.4°F)
Object temperature mode:
100°C (212.0°F)
The temperature measured
is lower than
Human thermometer mode:
34°C (93.2°F)
Object temperature mode:
0°C (32.0°F)
The operating temperature is
not in the range 16°C~40°C
(60.8°F~104°F)
Solution
Operate the thermometer only
between the specified temperature
ranges. If necessary, clean the
sensor tip. In the event of a
repeated error message, contact
your retailer or Customer Services.
Operate the thermometer only
between the specified temperature
ranges.
GB | 9

Publicidad

Tabla de contenido
loading

Productos relacionados para Salter TE-250-EU

Tabla de contenido