I PACK CONT ENTS
-
One assembled valve (1) including:
.
One valve seat (1.
.
One valve head (1.
-
One flat gasket (2)
-
This instruction sheet.
II USE
The 1S162 valve is intended for use in lower limb prosthesis only. Any other use is not permitted.
III INDICATIONS / CONT RA -INDICATIONS
No contra-indications for using the 1S162 valve in a lower limb prosthesis, whatever the patient weight and activity level.
IV MAKING T HE SOCKET
Make the socket as u sual and mark the position
of the valve on this socket, then
a) On the lankmark, drill a 8.5 mm Ø hole
perpendicular to the socket wall
b) Cut a M10 x 1.5 thread.
V PREPARING T HE VALVE
Disassemble the valve head (1.
a)
valve seat (1.
), then place the flat gasket (2)
1
on the valve seat
b)
Using a 5 mm hex wrench (3), screw the
valve seat (1.
) into the socket and take care
1
that the flat gasket (2) be compressed
enough.