Stereo microscope
Stemi 508
Before using the Stemi 508, read the detailed operating manual carefully and comply with the safety provisions contained in it.
Description
1
Eyepieces, adjustable
2
Stemi 508 microscope body
3
Camera port 60N with camera adapter
0.5x for C-mount cameras
(only Stemi 508 trino)
4
Zoom knob with switchable clickstops
Handle
5
6
Microscope mount
7
Stand
8
Focusing knob to focus the microscopic
image
9
Clamping screw to fasten the
microscope body in the stand
10
Rotary/push-button for switching and
dimming the reflected-light illuminator
11
Rotary/push-button for switching and
dimming the transmitted-light
illuminator
12
Spot illuminator K LED (optional)
13
Lever for adjusting the transmitted-
light illuminator (stand K EDU)
14
Insert plate for positioning the specimen
15
Detent spring for insert plate
16
RJ 12 socket on microscope mount to
connect reflected-light illuminators
(bottom)
17
Allen key, SW 3
18
Power connection socket
19
Power cable
20
ON/OFF switch
21
Eyepiece tube
22
Shifting the eyepiece tubes to adjust
the individual eye distance
23
(Individual) eye distance
24
Distance between eye and eyepiece
insight (approx. 2 cm)
435064-8044-008 e
Quick Start
Short operating manual
Installation
Remove Stemi 508 from the packing and
check for integrality according to the
delivery note.
• Place stand (7) on a level, firm support.
• Should the Stemi 508 body (2) not yet
have been mounted, insert it into the
opening of the microscope mount (6),
align it and fasten it securely using the
clamping screw (9) and the Allen key
(17).
• Insert the two eyepieces (1) into the
eyepiece tube (21) to their stop.
• To place the insert plate (14) in the
mount in the base of the stand, press
the detent spring (15), making it lock in
carefully.
• Insert power cable (19) into the power
connection socket (18) of the stand (7)
and connect it to a plug box.
• Switch the microscope ON or OFF via
the switch (20).
Safe disconnection from the power
supply is only ensured by removing
the power plug. The switch on the
microscope
only
standby mode.
Do not replace detachable power
cables by inadequately designed
power cables. Only the specified
power cables may be used.
Operation
The microscope is
switched ON.
1.
Position the specimen centrally on
the insert plate (14).
2.
Switch light ON; to this effect, press
rotary knob (10 or 11) briefly and
turn to adjust brightness.
3.
Set the two adjustable eyepieces (1)
to "zero" (white dot on "0").
4.
Set individual eye distance (23) by
shifting (22) the eyepiece sockets
(21) laterally, so that only one
circular image is visible on looking
into the eyepieces with both eyes.
To this effect, a distance of
approx. 2 cm must be maintained
between the eye and the eyepiece
insight (24).
5.
Set the lowest magnification using
the zoom knob (4).
6.
Focus on a small distinct detail in
the center of the specimen using
the focusing knob (8).
7.
Set the highest magnification using
switches
into
the zoom knob (4).
8.
Focus on the distinct specimen
detail again using the focusing
knob (8).
9.
Return to lowest magnification
using the zoom knob (4).
10. Correct any blur in the microscopic
image separately for each eye using
the adjustable eyepieces (correction
of individual visual defect).